          【Case Study】Nasal Irrigator Registration in China

          from CIRS by

          1. Introduction

          Product Name: nasal irrigator, nasal wash

          Product Composition: consists of a main unit, a water tank, a nasal wash (adult and child type), and a power adapter.

          Working Principle:

          After the nasal irrigator is energized, the motor converts the electrical energy into mechanical pressure. The pressure is used to deliver warm water/physiological saline into the nostrils, through the nasal vestibule, sinus, and nasal passages around the nasopharynx, and from the other side of the nostrils. Through the above path, the sterilizing action of the physiological saline itself and the impact force of the water flow are used to discharge the accumulated dirt and harmful substances in the nasal cavity, thereby achieving the purpose of cleaning and daily care of the nasal cavity.

          Typical Structure:

          Nasal Irrigator

          Adult type Child type

          No.

          Name

          Description

          1

          Head

          The outlet of the water flow controls the water volume stably and sprays evenly

          2

          Nozzle release button

          Use and pause control buttons

          3

          Host

          Control the operation of the nasal irrigator

          4

          Rotary switch

          Adjustment circuit closure

          5

          Water tank

          Water storage

          6

          Power adapter

          Connected to the power supply

          Scope of application: For nasal irrigation and daily nasal hygiene

          Applicable people: Suitable for adults and children with nasal cleansing needs

          2. Classification

          Classification Code: 14-07-01 Class II Flushing device

          Product Description: Usually composed of a host, a heating module, and a power module

          Expected Usage: For rinsing the natural cavity (excluding vaginal), rinsing the tissue during and after surgery, and warming the rinsing fluid

          3. Registration Unit

          1. Different principle

          Passive polymer material nasal irrigators and active electric irrigators cannot be registered together.

          2. Sterile and non-sterile

          The polymer material type irrigator is a class II device provided by sterilizing, and a class I device provided by non-sterile; the electric nasal lavage device is class II.

          3. Whether it contains drugs

          Separate registration with medicinal ingredients and without drugs.

          4. Technical Requirements

          1. Appearance

          The appearance of the nasal irrigator should be neat, smooth and uniform in color; no defects such as sharp edges, burrs, cracks, etc.; the characters and signs on the panel should be clearly visible; the installation and connection parts should be tight, the movable parts should be used, the rotation is flexible and reliable; the water tank is leak-free. .

          2. Tank capacity

          The water tank capacity is 700ml.

          3. Flushing pressure

          The water pressure at the outlet of the nasal wash should be between 0.1 MPa and 0.3 MPa.

          4. Flushing flow rate

          The water flow rate at the outlet of the nasal wash should be no more than 10 ml/s.

          5. Flush height

          The outlet height of the adult nasal wash outlet is 4.0 cm ± 1.0 cm; the height of the child's nasal wash outlet is 2.0 cm ± 1.0 cm.

          6. Irrigator piping system effluent chemical properties

          6.1 pH

          The difference between the pH of the eluate and the blank control solution should be ≤1.0;

          6.2 Heavy metal content

          The heavy metal content of the eluate should be ≤1.0 μg/mL.

          7. Electrical safety requirements

          Should meet the requirements of GB9706.1-2007.

          8. Electromagnetic compatibility requirements

          Should meet the requirements specified in YY 0505-2012.

          9. Environmental test requirements

          It should meet the requirements of the climatic environment test group II and the mechanical environment group II in GB/T 14710-2009.

          Table 1 Relevant Product Standards

          Standard Number

          Standard Name

          GB/T 191-2008

          Packaging storage and transportation icon

          YY/T 0316-2016

          Medical device risk management for medical devices

          YY/T 0466.1-2009

          Medical devices - Symbols for labeling, marking and providing information on medical devices - Part 1: General requirements

          GB/T 14710-2009

          Medical electrical environment requirements and test methods

          GB 9706.1-2007

          Medical electrical equipment - Part 1: General requirements for safety

          YY 0505-2012

          Medical electrical equipment - Part 1-2: Safety commons Parallel standard: Electromagnetic compatibility Requirements and testing

          GB/T 14233.1-2008

          Medical infusion, blood transfusion, injectables - Test methods - Part 1

          YY(/T): Industry standard (Recommendation)

          GB(/T): National standard (Recommendation)

          5. Clinical Trials

          Clinical Trials: required

          6. Registration Cost and Duration

          1) Product Testing

          Test items

          Testing duration

          Testing fee(USD)

          Safety Performance

          60 working days

          3,000

          EMC (Internal and external circuits )

          90 working days

          4,500*2

          Biocompatibility

          60 working days

          3,500

          2) Technology Approval

          60 working days

          Another 60 working days if the 1-year supplement is required

          3) Administrative Approval

          20 working days

          If you would like to get the detailed registration proposal, please contact us via md@cirs-group.com.

            

          We have launched a LinkedIn newsletter to keep you up to date on the latest developments across the chemical industry including food and FCMs and personal and home care.

          Contact Us
          +86 571 8720 6559 (GMT+8 8:30~17:00)
          Contact Us
          +86 571 8720 6559 (GMT+8 8:30~17:00)
          1. <dfn id="8ngdu"></dfn>

            1. <dfn id="8ngdu"></dfn>
              <rp id="8ngdu"><address id="8ngdu"></address></rp>
                <video id="8ngdu"><menu id="8ngdu"><kbd id="8ngdu"></kbd></menu></video>

                1. <b id="8ngdu"></b>

                亚洲国产精品激情在线观看